32
OPERATING PROCEDURES
:
Tracing
{
Semiauto tracing
This is the procedure to fit the stylus into the groove by hand before tracing for the frame whose stylus
is not automatically set in the groove.
This is for frames, especially plastic ones, whose grooves are not in the middle of the edges.
1
Set frames into the tracing unit.
See Step 1 of “Both-eye tracing” (Page 30).
2
Press the Both-eye tracing button
for three seconds.
The stylus moves and pauses in the start position of tracing. The
button blinks.
In single-eye tracing, press the Right-eye tracing button
or Left-eye tracing button
for three seconds in stead of the
button.
3
Fit the stylus into the groove by hand.
4
Start tracing.
Press the
Both-eye tracing button
.
After tracing one frame, the stylus moves to another frame and pauses in the tracing start
position.
5
Trace the other frame in the same manner as Steps 3 and 4.
Содержание ICE-1000
Страница 1: ...INTELLIGENT BLOCKER Model ICE 1000 OPERATOR S MANUAL...
Страница 8: ...VI...
Страница 114: ...104 SPECIFICATIONS AND ACCESSORIES Standard Configuration 5 3 2 Optional accessories Barcode scanner RS 232C cable...