BEFORE USE
:
Before First Use
1
1.6.1
Tracing unit calibration
Perform calibration of the tracing unit using the supplied standard frame and standard pattern.
This calibration will be performed by service persons at the time of installation.
It is recommended to perform the tracking unit calibration to ensure high tracing accuracy before trac-
ing frames or patterns. In addition, when a measured circumference involves an error, perform the
tracing unit calibration.
1
Perform calibration with the standard frame.
1) Turn on the ICE-1000 and input a desired JOB# on a layout display.
2) Set the supplied standard frame (
45) to
the tracing unit.
Set the standard frame with the surface with
the inscribed numbers up so that the rim clips
are set over eight semicircular hollows on both
surfaces of the frame.
3) Press the
button while depressing
the
button.
The lamp of the
button blinks.
4)
Press the
button.
The calibration is performed.
5) After the calibration is complete, start tracing the standard frame with the
button.
6) Press the Tracer button.
7) Check the traced data (circumference) on
the layout display.
When it is not within 162.83 ± 0.05 mm, per-
form calibration with the standard frame again.
8) Remove the standard frame.
Standard frame
For calibration of frame tracing by the stylus
Standard pattern
For calibration of pattern tracing by the pattern tracing pin
Hollows
Inscribed numbers
Circumference of the traced outline