24
BEFORE USE
:
Before First Use
2
Perform calibration with the standard pattern.
1) Set the supplied standard pattern into the
tracing unit.
Fit the standard pattern with the surface with
the inscribed “A” facing the Pattern setting unit.
See Steps 1 to 2 of 2.4.2 Pattern tracing for the
method of setting a pattern.
2) Press the
button while depressing
the
button.
The lamp of the
button blinks.
3) Press the
button.
The calibration is performed.
4) After the calibration is complete, start tracing the standard pattern with the
button (or
the
button).
5) Press the Tracer button.
6) Check the traced data (circumference) on
the layout display.
When it is not within 162.83±0.10 mm, perform
calibration with the standard pattern again.
7) Remove the standard pattern.
Standard pattern
Pattern setting unit
Circumference of the traced outline