90
CONNECTION AND SETTINGS
:
Setting Parameters
3
Press the Screen change tab to switch to the
Parameter 2 screen or Parameter 3 screen
and change the setting contents in the same
way.
4
Press the Exit button to return
to the layout screen.
[Parameter description: Parameter 1 screen]
1) System: Mini LAB, Blocker/LAB, Entry/LAB, Extended/LAB, Blocker/LE, Blocker/VCA
Factory setting: Mini LAB
This parameter sets the system configuration where the ICE-1000 is connected.
Blocker/LAB and Blocker/LE should be selected when the ICE-1000NT (with no tracing unit
equipped) is used.
Blocker/VCA should be selected for special models.
2) ID No.: 0 to 9999
Factory setting: 32
This parameter sets the ID No. for multiple ICE-1000 units in a network. Set any number
from 0 to 9999 as an identification number.
3) Server No.: 0 to 9999
Factory setting: 32
This parameter sets the ID No. for the host device that is the server on a network. Set any
number from 0 to 9999.
4) Baud rate: 38400, 9600
Factory setting: 38400
This parameter sets the communication speed (Baud rate) for data communication using
EDGER 1 and EDGER 2 ports.
• Quickly press the Exit button.
Depressing the button displays the
System 1 screen.
When the System 1 screen is dis-
played, press the Exit button quickly.
The screen returns to the layout dis-
play.
Do not change the parameter settings
on the System 1 screen as they are
provided for compensation purposes.