AC Modes
5
Keysight 5500 SPM User’s Guide
5-5
Figure 6-3
ACAFM tab in AC Tune window.
7
Set the Drive (%) to
10 %
. This is the initial amplitude of the AC
drive signal, stated as a percentage (0-100 %) of the maximum
available 10 V.
8
In the Auto Tune tab (
) of the AC Tune
window, enter the
Start
and
End
frequencies (in kHz) for the tuning sweep. For a new
or unknown cantilever, use the stated minimum and maximum
frequencies given on the storage box. If the resonance frequency
range is known more exactly, a smaller range can be used to speed
the tuning process.
NOTE
If the system is using the AC source in the HEB, the Setpoint needs to
be set to zero to perform Auto Tune. If the system has an external
AC/MAC Mode Module, the Setpoint can be any value during Auto
Tune.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...