Scanner Maintenance and Calibration
5
Keysight 5500 SPM User’s Guide
5-6
hysteresis is that the trace will be offset from the retrace, as in
Figure 8-2
Scanner hysteresis before correction. The yellow line is the
Trace image, and the blue is the Retrace.
Other Characteristics
Bow
During raster scanning, the free end of the scanner moves in an arc over
the full range of the scanner, as opposed to a flat line in a plane above
the sample. Bowing is minimized by the “balanced pendulum” design of
the Keysight scanners. Residual bowing is typically accounted for by
“flattening” algorithms in the PicoView software.
Cross Coupling
Cross coupling is the effect in which movement of the scanner along
one axis (usually X or Y) causes unwanted movement along the other
axis (Z). Systems using tube scanners are more susceptible to geometric
cross coupling because a single four-quadrant tube provides motion in
all the three axes. The larger Keysight multi-purpose scanners, with 90
micron X/Y scan range, use separate piezoelectric elements for X/Y
movement and for Z movement. This configuration helps to reduce the
cross coupling between different axes. Smaller range scanners (e.g., 10
micron X/Y scan range) use two sets of plates (one each for X and Y
movement) and a single tube for Z motion. Since the range is small with
these scanners, the effect of cross-coupling is minimal.
Aging
Aging is a time-dependent effect in which the sensitivity (extension per
unit of field) of the scanner decreases, approximately exponentially,
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...