Closed-Loop Scanners
5
Keysight 5500 SPM User’s Guide
5-14
The goal of the calibration procedure is to make the Z Sensor output
appear like that shown in
Figure 9-7
Expected output for Z sensor calibration
The plot shows the output of the Z sensor as a function of Z piezo
displacement. The Z sensor output ranges from -10 V to +10 V over the
entire range of the Z piezo. This procedure sets the range just slightly
under ± 10 V.
13
In the Spectroscopy plot window, turn Toggle Auto Full Scale
Off.
14
Set Y Min to
-9.8
and Y
Max to
+9.8
.
15
In the Z Sensor
section of the Scanner Setup dialog box, adjust the
Offset
to change the vertical position of the lines until their left edge
is as close to -9.8 V as possible.
16
Adjust the
Gain
to change the slope of the lines until they meet in
the upper right corner.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...