Closed-Loop Scanners
5
Keysight 5500 SPM User’s Guide
5-2
defined height above the sample for specialized experiments, such as
measuring how long molecules stretch and relax (
Figure 9-1
Setting Z position in the Position tab in the Servo dialog box
X/Y/Z Closed-Loop Scanner
The X/Y/Z closed-loop scanner includes the functionality of the Z-axis
scanner described above, with additional encoders on the X and Y axes.
These sensors allow for very linear scans and also make it easy to move
to precise locations on the sample.
Calibration
Keysight recommends calibrating the Gains, Offset and Sensitivity
settings in the Closed Loop tab of the Scanner Setup dialog box prior to
each use of the closed-loop scanner. Begin by calibrating the scanner
following the open-loop instructions in the
For a Z-axis scanner, skip to
NOTE
The procedures and descriptions in this User Guide are for PicoView
1.12x and higher.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...