MAC III
5
Keysight 5500 SPM User’s Guide
5-30
the check box to replace the Ref Set value with the
lock-in signal.
Select the
Sum
check box plus
GND
to pass the Ref
Set value from the AFM controller directly to the
microscope. These are the default settings.
BNC 1 and 2
The MAC III controller includes seven output
channels: Deflection, Friction, SP and AUX 1-4.
These output signals are routed to the AFM
controller. They can also be duplicated at the two
BNC connectors on the MAC III controller for
additional routing flexibility.
By default, Deflection is routed to BNC1, and
Friction is routed to BNC2. The actual signal sent to
the BNC connectors is selected in the Output
Channels, described next.
Outputs
The MAC III controller includes seven outputs
(Deflection, Friction, SP, AUX 1-4) that are routed
to the AFM controller for imaging.
Each output can carry one of thirteen signals: the
Amplitude, Phase, X Component or Y Component
of the three lock-in signals; or the output of the
MAC III internal servo. Selecting GND for any
output sets its output to 0.
By default, the Deflection output carries Amplitude
1 (the amplitude of Lock-in 1 output). The Friction
output carries Phase 1. The remaining outputs are
set to GND.
Select the
Pass Through
check box for each output
to pass the signal directly from the microscope to the
AFM controller without further contribution from
the MAC III controller. By default, Pass Through is
selected for each output.
NOTE
In KFM Mode, the SP (Surface Potential) channel is set to Servo Output
and the
Pass Through
check box is cleared. The Servo Output is the DC
bias produced by the servo to counteract the sample bias.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...