MAC III
5
Keysight 5500 SPM User’s Guide
5-15
allow the correct signals to pass through to the Laser Alignment
window.
STM
STM does not require any MAC III-specific controls.
LFM
Contact mode does not require any MAC III-specific controls.
DLFM
In DLFM mode, Lock-in 1 is used to oscillate the tip parallel to the
sample surface, with the Friction channel as its input. The Drive
Mechanism is set to
AAC
. Cantilever deflection is controlled as in
Contact Mode.
Choose
Mode > DLFM
, then choose
Controls > AC Mode
to open the
DLFM Controls window:
Figure 11-8
DLFM Controls window
CAUTION
The piezo element used to oscillate the cantilever in the DLFM nose
assembly is partially exposed; therefore, DLFM should
never
be used
for imaging in liquid.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...