Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-18
10
In the Servo window set the I Gain and P Gain to
5 %
.
11
In the Scan and Motor window’s
Scan
tab, enter:
a
Scan Speed of
1-2
ln/s.
b
Resolution of
256
.
c
Scan
Size
(in microns).
d
X Offset
and/or
Y Offset
values to set the center of the scan.
12
In the Realtime Images
window, display three images for
Topography
,
CSAFM/Aux BNC
(the Phase signal via the MAC
controller), and
HEB Aux
(the Amplitude signal via the HEB). If
using a MAC III controller, select
Aux 1
and
Aux 2
.
13
In the Scan and Motor window, click the
Up
or
Down
blue arrows to
initiate a scan.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...