Contact Mode Imaging
5
Keysight 5500 SPM User’s Guide
5-11
Gain settings of 5-10 % are typical, though some iteration will most
often be required to optimize the gains for a particular sample.
Scan Settings
In the Scan Controls section, the scan
Speed
and
Resolution
(Points/Line) will both affect image quality.
A faster scan speed decreases imaging time but may not allow the
system sufficient time to accurately track the topography. A typical scan
speed will be 2-5 lines/second for smooth surfaces. For rougher surfaces
a lower scan speed may be needed.
A typical resolution of 256 pixels/line provides good resolution and
speed. Increasing the resolution will improve image quality but will
require longer imaging times.
One good option is to scan a large region at low resolution and high
speed, and then to zoom in on a region of interest for a high resolution
scan at lower speed. After completing the large scan, use the Offset and
Size values in the Scan and Motor window to adjust the scan to cover
the region of interest. Increase the Resolution, decrease the Speed, and
re-scan the zoomed region.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...