background image

 

 

 

 

 

LSM 510 and LSM 510 META

Laser Scanning Microscopes 

 

Brief Operating Manual 

Release 3.5 

July 2005 

Содержание LSM 510 Inverted

Страница 1: ...LSM 510 and LSM 510 META Laser Scanning Microscopes Brief Operating Manual Release 3 5 July 2005...

Страница 2: ...e LSM 510 and LSM 510 META laser scanning microscope including its original accessories and compatible accessories from other manufacturers may only be used for the purposes and microscopy techniques...

Страница 3: ...nents switch also to ON Now the complete system is ready to be initialized with the LSM Software Switching on the HBO 100 mercury lamp Switch on the HBO 100 mercury lamp via the switch of the power su...

Страница 4: ...10 Switchboard window Clicking on this button activates the complete LSM hardware on line mode Click on the Start Expert Mode button in the LSM 510 Switchboard window The LSM 510 Expert Mode Main menu...

Страница 5: ...ate toolbar Click on the New button in the File subordinate toolbar The Create New Database window appears Select drive C or D from pull down menu Create a new directory if needed Fig 6 Create New Dat...

Страница 6: ...cular tube of the microscope Click on the VIS button to set the microscope for direct observation via the eyepieces of the binocular tube lasers are off Click on the TV button to set the microscope ca...

Страница 7: ...ting the microscope for reflected light Click on the Reflected Light button to open the shutter of the HBO 100 mercury lamp Click on the Reflector button and select the desired filter set by clicking...

Страница 8: ...ick on the Single Track button in the Configuration Control window Fig 12 Click on the Descanned button if necessary The Beam Path and Channel Assignment panel of the Configuration Control window disp...

Страница 9: ...he first line of the Configurations list box and click an Store For loading an existing configuration select it in the list box and click on Apply For deleting an existing configuration select it in t...

Страница 10: ...position and diameter When Line button is selected the same rules apply as for Frame Fast Settings Add Track button An additional track is added to the configuration list The maximum of four tracks c...

Страница 11: ...lts in the best signal to noise ratio Scan speed 8 usually produces good results Us speed 6 or 7 for superior images Choosing the dynamic range Select the dynamic range 8 or 12 Bit per pixel in the Pi...

Страница 12: ...same Optical Size This is important for colocalization studies Image acquisition Once you have set up your parameter as defined in the above section you can acquire a frame image of your specimen Use...

Страница 13: ...gh it appears blue on the screen Blue zero minimum Adjusting the laser intensity Set the Pinhole to 1 Airy Unit Fig 21 Set the Detector Gain high When the image is saturated reduce AOTF transmission i...

Страница 14: ...upper position of the specimen area where the Z Stack is to start Click on the Mark First button to set the upper position of the Z Stack Then focus on the lower specimen area where the recording of t...

Страница 15: ...ing off the system Click on the File button in the Main menu and then click on the Exit button to leave LSM 5 software program Fig 5 If any lasers are still running you should shut them off now in the...

Страница 16: ......

Отзывы: