Scanner Maintenance and Calibration
5
Keysight 5500 SPM User’s Guide
5-11
If the spacing is not identical for the two sets of features, adjust the X
Non-linearity terms in the Open Loop tab of the Scanner Setup dialog
box, according to the following equation:
where
StartSize
= size of features at the start of the scan
EndSize =
size of features at the end of the scan
CurrentTerm
= current non-linear correction term.
Adjust
Non-linearity 1
for the Trace scan and adjust
Non-linearity 2
using the Retrace scan.
You can also use the diagram in
Figure 8-7
Correction diagram for X non-linearity
After adjusting the X Non-linearity parameters, re-scan the calibration
target. The spacing between features should be approximately the same
across the scan range for the Trace and Retrace scans.
X Hysteresis
Place a vertical cursor at the same location in the Trace and Retrace
images. The cursor should cross the same features in both images. If this
NewTerm
StartSize CurrentTerm
EndSize
-----------------------------------------------------------------
=
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...