Keysight 5500 SPM User’s Guide
5-31
Other Tab
The Other tab includes additional parameters that control MAC III
operation:
Figure 11-14
Advanced AC Mode window: Other tab
Drive
Selects the drive mechanism (AAC, standard MAC
or Top MAC).
I Gain
The Integral Gain to the MAC III internal servo
loop. The default value is 0.
P Gain
The Proportional Gain to the MAC III servo. The
default value is 0.
Setpoint (V)
The voltage which the servo will try to maintain.
Input
Routes the Amplitude, Phase, X Component or Y
Component from the three lock-in signals to the
MAC III servo input. This is the signal that the servo
will maintain at the selected Setpoint. Selecting
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...