Closed-Loop Scanners
5
Keysight 5500 SPM User’s Guide
5-5
Figure 9-3
Advanced tab settings for CL calibration
9
Without a sample in place, use the laser alignment controls and
camera window to align the laser and detector for Contact AFM
mode.
10
In the Scan and Motor
window enter:
a
A very large number (e.g., 9999) for the Scan
Size. It will adjust
automatically to the maximum allowed value
.
b
Speed of
1-2 lines/second
.
c
Resolution of
256
.
11
In the Realtime Images window, create two image buffers with the
following settings:
•
Input set to
X Sensor
.
•
Flattening set to
None
.
•
Display Range
set to
-10 V to +10 V
.
•
For one image display
Trace
and for the other display
Retrace
.
•
Set both buffers to
Main
scan mode.
12
Choose
Tools > Realtime Cross Section
so the individual scan lines
will be displayed in the Cross Section window. Set the range to
±10 V.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...