Keysight 5500 SPM
User’s Guide
12
Liquid Cell
Liquid Cell with Standard Sample Plate
The liquid cell enables
in-situ
AFM or STM imaging for better control
under realistic environments. The cell is made from chemical-resistant
polycarbonate and can be used with a wide variety of liquids. The cell
can be used in conjunction with a standard, MAC Mode or temperature
control sample plates.
A flow-through version of the liquid cell is also available with 0.9 mm
holes included for tubing.
Eight-degree nose assemblies are recommended for imaging in liquid
because the smaller angle takes into account the different angle the laser
makes as it goes in and out of the fluid, compared to operation in air. A
Top MAC nose assembly is available with an eight-degree nose
assembly option.
shows the components of the liquid cell: two retaining clips,
an O-ring gasket and the liquid cell plate.
components as assembled on a standard sample plate. Note that, when
assembled, the sample itself comprises the bottom of the liquid
CAUTION
Some nose assemblies, such as the two-piece nose assemblies and
CSAFM nose assembly, are not sealed and should never be used for
imaging in liquid. Be sure to use only approved nose assemblies for
imaging with the liquid cell.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...