Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-37
Figure 7-20
Initial Servo settings for FlexGrid Spectroscopy
4
Choose
Controls >
Spectroscopy
or click the
Spectroscopy
toolbar button
to open the Spectroscopy
window.
5
Verify the Plot Type is set to
Force vs Distance
, either in the
Spectroscopy drop down menu
or in the status bar at the bottom
of the Spectroscopy window.
6
Approach
the tip.
7
In the Spectroscopy Basic tab,
set the Sweeps as needed to
either
1
or
.
8
In the Spectroscopy Setup tab¸
set the Duration to
1
.
9
Set the Data Points to
200
.
10
Maximize the Start and
minimize the End values
a
Enter a large number for the
Start value (ie 999) and
PicoView will adjust to the
maximum value for the scanner.
b
Enter a large negative number for the End value (ie. -999) and
PicoView will adjust to the minimum value for the scanner.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...