MAC III
5
Keysight 5500 SPM User’s Guide
5-27
to the drive signal. This option is used in
KFM Mode, in which the servo acts to
maintain a DC Tip Bias that counteracts any
electrostatic field on the sample. The servo
output, therefore, will change as the sample
charge changes; this value is also fed to the
SP channel for imaging.
Phase Shift (°)
Each lock-in includes two orthogonal
reference signals. This parameter will shift
the phase of the reference signals with
respect to the drive signal (they will remain
orthogonal to each other). This feature is
useful, for example, to maximize the real
component of the drive signal in KFM
mode. The default value is 0.
Phase Compensation
Phase Compensation adds a frequency
dependent phase offset to compensate for
any phase shifts in the electronics.
Sum External Drive
Select this check box to add a signal from
the ASAB
External Drive Input
to the
lock-in drive signal. By default the check
box cleared.
Y Component from AUX Select this check box to replace the Y signal
with the AUX input. This option essentially
converts the lock-in to an A/D converter,
providing a quick method for measuring an
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...