MAC Mode
5
Keysight 5500 SPM User’s Guide
5-4
page 12). When using Acoustic AC Mode, this connection is not
necessary.
Hardware and Sample Setup
Most hardware and sample setup options with MAC Mode are identical
to those for standard AAC and MAC Mode operation, as covered in
set up the microscope for imaging. Note the phase conversion
coefficient is 18 ° per volt for the MAC Mode controller.
In AAC Mode the drive signal can be provided by either the Head
Electronics Box or the MAC Mode controller. To use the MAC Mode
controller as the AC source, choose
Controls > Setup > Components
,
then select the
Serial Port AC Mode Controller
check box. The
system will now use the drive signal from the MAC Mode controller,
through the HEB to the microscope.
Upon startup, the software may instruct you to update the system’s
firmware. Follow the on-screen instructions to do so.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...