Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-22
Set up the Interleave parameters. In the Scan and Motor window, click
on the
Interleave
tab (
).
Figure 7-13
Interleave tab used for MFM imaging
1
Select the
Drive (%) Lock-in 1
check box.
2
Set the Lock-in 1 Frequency to
0.100 kHz
above the cantilever
resonance frequency. In the case above (
resonance frequency was 72.539 kHz, the Interleave Frequency
would be set to 72.639 kHz.
3
Set the Interleave Drive (%) so the amplitude is about 1/5 the free
amplitude used for the topography scan:
a
Choose
Controls > AC Mode
to open the ACAFM Controls
dialog box.
b
Adjust the
Drive (%)
to determine the value needed to reduce the
amplitude for the Interleave setting of 1/5 free amplitude.
c
Enter the Drive (%) value from the ACAFM Controls dialog box
into the Interleave Drive (%) (
) for MFM imaging.
d
Re-set the Drive (%) in the the ACAFM Controls dialog box to
its original value.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...