Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-7
). Connect the electrode to the sample,
ensuring good contact.
Figure 7-7
Sample on plate with electrode attached
6
Place the sample plate on the microscope.
7
Plug the 3-pin EC connector of the EC/MAC cable into the 3-pin
socket on the sample plate. Plug the other end of the cable into the
EC/MAC socket on the microscope.
8
In PicoView, choose
Mode > STM
.
9
In the Servo
window enter the Bias Voltage (
). Typical
values are
50-200 millivolts
(0.05-0.200 V). A positive bias
indicates current flow
from
the tip
to
the sample, and vice versa for
negative bias.
NOTE
The sample plate cable can transfer low levels of vibration to the
sample. During very high resolution imaging this can affect images
quality. We recommend first plugging the sample plate cable to the
flexible 3-wire umbilical included with the sample plate. The umbilical
should then be plugged in to the microscope base. The umbilical’s
individual wires tend to reduce the transfer of vibration.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...