Additional Imaging Modes
5
Keysight 5500 SPM User’s Guide
5-33
3
Choose
Controls > Spectroscopy
or click the
Spectroscopy
toolbar
button
to open the Spectroscopy window.
4
Verify the Plot Type is set to
Force vs Distance
, either in the
Spectroscopy drop down menu
or in the status bar at the bottom
of the Spectroscopy window.
5
Approach
the tip.
6
In the Spectroscopy Basic tab,
set the Sweeps as needed to
either 1 or
.
7
In the Spectroscopy Setup tab,
set the Duration to
1
.
8
Set the Data Points to
200
.
9
Maximize the Start and
minimize the End values:
a
Enter a large number for the
Start value (ie 999) and
PicoView will adjust to the
maximum value for the
scanner.
b
Enter a large negative
number for the End value (ie.
-999) and PicoView will
adjust to the minimum value for the scanner.
10
If using a calibrated CL scanner, select the
CL Sweep
check box (as
) in the Z Closed Loop section of the Setup tab.
Figure 7-18
Z closed loop section of Spectroscopy Setup tab with
CL
Sweep
check box selected
To save successive force measurements, have files saved automatically
after each force curve:
1
In the Spectroscopy window, choose
File > Set filename ...
.
2
In the
Save As
dialog box, enter a filename for the data set. Note the
file type will default to *.mi.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...