MAC III
5
Keysight 5500 SPM User’s Guide
5-20
10 V, beyond which the signal will be clipped. The
default value is x1 (the amplitude times 1).
Zero Phase
Sets the phase at the current frequency to zero,
making it easier to interpret phase changes from the
current value.
Q Control On
By applying a phase-shifted version of the
cantilever deflection signal on top of the drive
signal, Q control can either increase or decrease the
effective quality factor of the system. Select this
On
check box to enable the Q Control feedback loop.
By default, Q Control is turned Off.
Drive (%)
Sets the amplitude of the Q Control phase-shifted
signal, stated as a percentage (0-100 %) of the
maximum available.
Optimize
Sets the optimal Q-Control Phase Shift and Drive.
The EFM tab shows the parameters for Lock-in 2. Drive, Frequency and
Gain settings have the same functions as described for Lock-in 1 in
As mentioned, Lock-in 2 is used as the source for the AC tip bias. You
will typically need to sweep the frequency of Lock-in 2 to ensure that
the electrical response of the cantilever does not interfere with the
mechanical response provided by Lock-in 1 and to see if there are any
other resonances present. To do so, select the
EFM Tune
check box,
then choose
Manual Tune
in the AC Tune window. Determining the
best frequency for your sample and tip will require some iteration. Two
rules typically apply:
•
The frequency should not be an integral factor of the Lock-in 1
frequency.
•
The frequency should not be close (within 10-20 kHz) to the
Lock-in 1 frequency.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...