AC Modes
5
Keysight 5500 SPM User’s Guide
5-6
Figure 6-4
Auto Tune parameters tab
9
Set the Peak Amplitude, which is the desired amplitude of cantilever
oscillation at the resonant frequency. A setting of
4
volts is a typical
starting value.
10
Enter an Off Peak value to offset the oscillation frequency from the
cantilever’s resonance frequency. A typical starting value is
-0.120 kHz
.
A negative Off Peak parameter is used to ensure good engagement of
the tip with the sample as a drive frequency below the resonant
frequency means the amplitude will slightly increase as the tip
approaches the sample.
11
Click the
Auto Tune
button.
The system will sweep several times through the range of frequencies
specified in the Auto Tune tab and will locate the resonant frequency of
the cantilever. The Drive (%) will be adjusted automatically to set the
amplitude at the resonant frequency to the value set in the Peak
Amplitude box. Finally, the AC drive frequency will be set to the
resonant frequency, plus or minus the value set in the Offset box.
shows the result of the Auto Tune process.
12
Focus the cantilever in the video window.
13
Rotate the video system focus knob very slightly towards you such
that the tip goes just out of focus (the focal plane is just below the tip
now).
14
Press the
Close
switch to raise the sample until the sample comes
into focus.
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...