AC Modes
5
Keysight 5500 SPM User’s Guide
5-4
5
Choose
Controls > AC Mode Tune
or click the AC Mode Tune
button in the toolbar (shown to the right) to open the AC Tune
window (
Figure 6-2
AC Tune window
The next step is to tune the oscillation signal to match the frequency of
the particular cantilever. The controls in the AC Tune window will be
used to configure the sweep through a range of frequencies. The
resultant plot should show one strong, sharp resonance peak. The
cantilever storage box or data sheet will indicate the range in which the
primary resonance frequency of the cantilever will be found.
There are two options for finding the resonant frequency of the
cantilever, Auto or Manual Tune. In most cases the Auto Tune
procedure will be used and is the procedure that will be discussed here.
6
In the ACAFM tab of the AC Tune window, set the Drive
Mechanism to AAC. See
Содержание 5500
Страница 1: ...Keysight 5500 Scanning Probe Microscope User s Guide ...
Страница 2: ......
Страница 9: ...Read This First N9410 90001 Keysight 5500 SPM User s Guide ix Declaration of Conformity ...
Страница 174: ...Additional Imaging Modes 5 Keysight 5500 SPM User s Guide 5 40 Figure 7 23 Plug in Images window ...
Страница 215: ...Closed Loop Scanners 5 Keysight 5500 SPM User s Guide 5 17 Save the calibration file X Y Z calibration is now complete ...
Страница 274: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 8 Figure 13 10 7500 9500 hot sample plate wiring diagram ...
Страница 275: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 9 Figure 13 11 5500 hot MAC sample plate wiring diagram ...
Страница 276: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 10 Figure 13 12 7500 9500 hot MAC sample plate wiring diagram ...
Страница 284: ...Temperature Control 5 Keysight 5500 SPM User s Guide 5 18 Figure 13 21 5500 Peltier Cold sample plate wiring diagram ...
Страница 297: ...Thermal K 5 Keysight 5500 SPM User s Guide 5 9 Figure 14 8 Select units after calibrating the Force Constant ...
Страница 330: ...Keysight Technologies 5500 SPM User s Guide Part Number N9410 90001 Revision H Keysight Technologies 2015 ...