219
About Measurement Times and Measurement Speed
D: Screen display time
Screen display
Screen display time
No display
0.0 ms
Display
MAX 0.3 ms
E: Memory save time
Memory function
Memory save time
ON, or IN
MAX 0.4 ms
OFF
0.0 ms
Wait time
(1) When the measurement frequency is changed
When the measurement frequency is changed, the wait time is 1.5 ms.
However, the measurement frequency is divided into the following eight frequency ranges. When
changing the frequency such that the old frequency was in one of these ranges and the new
frequency is in another, a wait of 2 ms is added.
Frequency range
4 Hz to 99.99 Hz
100.00 Hz to 1.0000 kHz
1.0001 kHz to 5.0000 kHz
5.0001 kHz to 10.000 kHz
10.001 kHz to 100.00 kHz
100.01 kHz to 1.0000 MHz
1.0001 MHz to 2.0000 MHz
2.0001 MHz to 8.0000 MHz
(2) When the measurement range or low Z high accuracy mode is changed
When changing the measurement range or switching low Z high accuracy mode on or off, a wait of
4 ms is added.
(3) When the measurement signal level is changed
When changing the AC measurement signal level, a 4 ms wait is added.
(4) When the DC bias is changed
A 1.5 ms wait is inserted when switching the DC bias on or off.
In addition, when changing the applied DC bias value, a 1 ms wait is added.
(5) When the cable length is changed
When changing the cable length between 0 m and 1 m (or 2 m or 4 m), a 2.5 ms wait is added.
(6) When switching to DC measurement (DC resistance measurement)
A 3.5 ms wait is inserted when switching from AC measurement to DC measurement.
(7) When changing multiple measurement conditions simultaneously
When changing multiple measurement conditions simultaneously, a wait of up to 6.5 ms is added.
10
Specifications
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...