173
Example Measurement Timing (Timing Charts)
•
The shorter the measurement time, the shorter the time that INDEX
―
―――
―
and EOM
―――
are HI (off).
The HI (OFF) time when receiving the INDEX
―
―――
―
and EOM
―――
signals may be too short depending on the input
circuit. By setting the EOM output method to
PULSE
, an adequately long HI (OFF) time can be configured.
The instrument can be configured to maintain the low (on) state for a preset time once
EOM
―――
changes to
LO (on) before reverting the signal to HI (off) after the completion of measurement.
When trigger input is received at EOM
―――
: LO (on) and INDEX
―
―――
―
: LO (on), the signal transitions to HI (off) when
measurement starts.
Setting the INDEX
―
――――
and EOM
―――
output method
Setting on the instrument:
See"Setting the EOM output method and output time" (p. 183).
Setting with a communications command:
See LCR Application Disc - Communications Command (
:IO:EOM:MODE
).
Setting the pulse width for which LO (on)
EOM
―――
is held
Setting on the instrument:
See "Setting the EOM output method and output time" (p. 183).
Setting with a communications command:
See LCR Application Disc - Communications Command (
:IO:EOM:PULSe
).
(Start-of-measurement signal)
(Analog end of measurement signal)
When HOLD
(Analog end of measurement signal)
When PULSE
(End of measurement signal)
When HOLD
(End of measurement signal)
When PULSE
T: pulse width
ON
In the above figure, “ON” indicates signal input and output, while “OFF” indicates the absence of signal
input or output.
(ON: HI, OFF:LO)
BCD signal timing
BCD single DM-n transition time based on the C1 and C2 signals
0.2ms max
0.2ms max
No. 1 Parameter
High-order digit
D1-0
to
D-4-3
BCDm-n
Low-order digit
(ON
High-order digit
)
(OFF)
Low-order digit
(ON)
C1
C2
High-order digit
(OFF)
No. 1 Parameter
(OFF)
No. 3 Parameter
(OFF)
0.2ms max
0.2ms max
No. 1 Parameter
High-order digit
D1-0
to
D-4-3
No. 2 Parameter
High-order digit
D1-0
to
D-4-3
No. 1 Parameter
Low-order digit
D1-0
to
D-4-3
No. 2 Parameter
Low-order digit
D1-0
to
D-4-3
9
External Control
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...