163
External Input/Output Connector and Signals
Pin I/O
*1
Signal name
Function
Logic
1
IN
TRIG
―
―
―
―
External trigger
(See "Input (IN) signal function details" (p. 167).)
Rising/
falling
Edge
2
-
(Unused)
-
-
-
3
-
(Unused)
-
-
-
4
IN
LD1
―
―
―
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative
Level
5
IN
LD3
―
―
―
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative
Level
6
IN
LD5
―
―
―
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative
Level
7
IN
C1
―
―
Switches between the high-order digit and low-order digit
during BCD output.
Negative
Level
8
-
ISO_5V
Isolated 5 V power output
-
-
9
-
ISO_COM
Isolated common signal ground
-
-
10
OUT ERR
―――
Outputs when a sampling error, overcurrent error, contact
error, Hihg-Z reject error, constant voltage/constant current
error, or voltage/current limit value exceeded error occurs.
Negative
Level
11
*2
OUT
BIN1
―
―
―
―
Generates output when the BIN measurement result is BIN1.
Negative
Level
PARA1-HI
―
―――
――
―
Generates output when the comparator judgment result is
HI for the No. 1 parameter.
D1-0
―
――
BCD output signal
12
*2
OUT
BIN3
―
―
―
―
Generates output when the BIN measurement result is BIN3.
Negative
Level
PARA1-LO
―
―――
――
―
Generates output when the comparator judgment result is
LO for the No. 1 parameter.
D1-2
―
――
BCD output signal
13
*2
OUT
BIN5
―
―
―
―
Generates output when the BIN measurement result is BIN5.
Negative
Level
PARA3-IN
―
―――
――
―
Generates output when the comparator judgment result is
IN for the No. 3 parameter.
D2-0
―
――
BCD output signal
14
*2
OUT
BIN7
―
―
―
―
Generates output when the BIN measurement result is BIN7.
Negative
Level
AND
―――
Outputs judgment results for two parameter measured val
-
ues (the result of a logical AND operation).
Generates output when both judgment results are IN. Ad
-
ditionally, generates output when either of the No. 1 or No. 3
parameters is IN and the other is undetermined.
D2-2
―
――
BCD output signal
15
OUT
BIN9
―
―
―
―
Generates output when the BIN measurement result is BIN9.
Negative
Level
D3-0
―
――
BCD output signal
16
OUT D3-2
―
――
BCD output signal
Negative
Level
17
OUT D4-0
―
――
BCD output signal
Negative
Level
18
OUT D4-2
―
――
BCD output signal
Negative
Level
19
OUT
OUT_OF_BINS
――
― ―― ―
―
――
Generates output when
the BIN judgment result does not match
any BIN.
Negative
Level
20
IN
C2
―
―
Switches between the No. 1 parameter and No. 3 parameter
during BCD output.
Negative
Level
21
IN
CALIB
――
―
―
―
DC adjustment request
(See "Input (IN) signal function details" (p. 167))
Negative
Level
22
IN
LD0
―
―
―
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative
Level
9
External Control
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...