204
Function specifications
•
USB
Terminator
CR+LF, CR (Default setting: CR+LF)
•
LAN
IP address
Four 3-digit values (from 0 to 255 each)
(Default setting: 192.168.000.001)
Subnet mask
Four 3-digit values (from 0 to 255 each)
(Default setting: 255.255.255.000)
Default gateway
Four 3-digit values (from 0 to 255 each) (Default setting: OFF)
Port number
1024 to 65535 (Default setting: 3500)
Terminator
CR+LF, CR (Default setting: CR+LF)
Instrument information
Serial number information
Displays the serial number.
Version information
Displays the software version and the FPGA version.
MAC address
Displays the MAC address.
USB ID
Displays the USB ID.
Self-check function
Panel test
Allows you to check for touch panel errors.
Panel calibration
Allows you to calibrate the touch panel.
Display status test
Allows you to check the screen display status and LED status.
ROM/RAM test
Allows you to check the instrument’s internal memory (ROM and RAM) for
errors.
EXT I/O test
Allows you to verify that output signals are output normally from EXT I/O
and that input signals can be read properly.
Time setting
Time and date setting
Sets the year, month, day, hour, minute, and second.
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...