193
Accessories and Options
10.3 Accessories and Options
Accessories: Refer to "Accessories" (p. 1).
Options: Refer to "Options (reference: open and short correction states)" (p. 2).
10.4
Function specifications
DC measurement (DC resistance measurement)
DC adjustment
Function
Functionality for measuring and canceling the circuit offset
ON/OFF selectable
When ON : Acquires the offset value at each measurement.
When OFF: Uses the offset value acquired initially in subsequent measurements.
Offset value acquisition
methods
(When set to “off”)
•
Change the Rdc range (the offset is acquired automatically when the range
is changed).
•
Change the Rdc low Z high accuracy mode setting (ON/OFF) while using a
range from 100 m
Ω
to 10
Ω
(value acquisition will be performed automati
-
cally when the setting is changed).
•
Change the adjust delay time (value acquisition will be performed auto
-
matically when the delay time is changed).
•Touch the
GET DCR OFFSET
key on the screen.
•
Input a signal from an external device to the instrument’s EXT I/O connector.
•
Send a communications command from an external device to the instru
-
ment’s interface.
Default setting
ON
DC delay
Function
Sets the time until DC resistance measurement starts.
Setting range
0.0000 s to 9.9999 s (0.0001 s resolution)
Default setting
0.0000 s
Adjustment delay
Function
Sets the time to wait before starting offset measurement so that a stable
measured value can be obtained.
Setting range
0.0030 s to 9.9999 s (0.0001 s resolution)
Default setting
0.0030 s
Power supply frequency setting
Function
Sets the frequency of the power supply being used.
Setting value
50 Hz/ 60 Hz
Default setting
60 Hz
Average
Function
Performs processing to average measured values.
Setting range
1 to 256 (1 step)
Averaging method
Arithmetic average
Default setting
1 (Average OFF)
10
Specifications
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...