115
Load Correction (Correcting Values to Match Reference Values)
Setting the correction frequency
1
Touch the
FREQ
key.
2
Enter the correction frequency with the
numeric keypad, and touch the unit key
to confirm the setting.
2
Touch when you
wish to perform load
correction during DC
measurement.
1
Touch when you wish
to cancel input. (This
dialog box will be
closed.)
Settable range: DC, 4 Hz to 8 MHz
*
*: The maximum frequency varies with the
If you make a mistake, touch the
C
key to
reenter the value.
3
Touch the
EXIT
key.
The dialog box will be closed.
Selecting the correction range
1
Touch the
RANGE
key.
2
Select the range for correction.
Touch when you wish to enable LOW Z op-
eration.
3
Touch the
EXIT
key.
The dialog box will be closed.
The selection of available ranges will vary with the correction frequency. For more information, see p. 211
of "10.6 Measurement Range and Accuracy".
5
Error Correction