107
Short Correction
Spot correction
Acquire the correction values at the set measurement frequencies. Measurement frequencies can
be set for up to five points.
Screen display method (For more information, see p. 27):
(Measurement screen)
ADJ
key>(
ADJ
screen)
SHORT
key
1
Touch the
ADJUST
key.
2
Touch the
SPOT
key, and then touch the
EXIT
key.
1
2
3
Select the correction point you want to
set or edit with the
key, and touch
the
EDIT
key.
2
1
Touch when you wish to cancel correction.
(The instrument will return to the screen shown
in step 2)
4
Enter a frequency for correction, and
touch the
SET
key to confirm it.
1
2
Until one of these keys is pressed for input
of a numerical value, the previous frequency
for which Spot correction was performed is
displayed.
•
Settable range: DC, 4 Hz to 8 MHz
*
*: The maximum frequency varies with the
• Touch the
C
key to cancel input.
•
If you attempt to set a measurement fre
-
quency greater than the maximum fre
-
quency for each cable length setting, it will
automatically be reduced to the maximum
frequency for each cable length setting.
•
If you attempt to set a measurement fre
-
quency lower than 4 Hz, it will automatically
be increased to 4 Hz.
The correction values from last time are dis
-
played in a confirmation screen.
Check that the measurement cable is in an
short circuit state.
5
Error Correction
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...