External Input/Output Connector and Signals
Pin I/O
*1
Signal name
Function
Logic
23
IN
LD2
―
―
―
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative
Level
24
IN
LD4
―
―
―
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative
Level
25
IN
LD6
―
―
―
Select panel number
(See "Input (IN) signal function details" (p. 167).)
Negative
Level
26
IN
LD_VALID
―
― ―
―
―
―
―
Execute panel load
(See "Input (IN) signal function details" (p. 167).)
Negative
Level
27
-
ISO_COM
Isolated common signal ground
-
-
28
OUT EOM
―――
This signal indicates that measurement is complete. “HI
(OFF)” indicates that measurement is in progress, while “LO
(ON)” indicates that measurement is complete. When LO
(ON), the comparator judgment results have been finalized.
Falling
Edge
29
OUT
INDEX
―
―――
―
Signal indicating that A/D conversion for the measurement
circuit has completed: When this signal changes from HI
(OFF) to LO (ON), the sample may be changed.
Falling
Edge
30
*2
OUT
BIN2
―
―
―
―
Generates output when the BIN measurement result is BIN2.
Negative
Level
PARA1-IN
―
―――
――
―
Generates output when the comparator judgment result is
IN for the No. 1 parameter.
D1-1
―
――
BCD output signal
31
*2
OUT
BIN4
―
―
―
―
Generates output when the BIN measurement result is BIN4.
Negative
Level
PARA3-HI
―
―――
――
―
Generates output when the comparator judgment result is
HI for the No. 3 parameter.
D1-3
―
――
BCD output signal
32
*2
OUT
BIN6
―
―
―
―
Generates output when the BIN measurement result is BIN6.
Negative
Level
PARA3-LO
―
―――
――
―
Generates output when the comparator judgment result is
LO for the No. 3 parameter.
D2-1
―
――
BCD output signal
33
OUT
BIN8
―
―
―
―
Generates output when the BIN measurement result is BIN8.
Negative
Level
D2-3
―
――
BCD output signal
34
OUT
BIN10
―
―
―
―
―
Generates output when the BIN measurement result is BIN10.
Negative
Level
D3-1
―
――
BCD output signal
35
OUT D3-3
―
――
BCD output signal
Negative
Level
36
OUT D4-1
―
――
BCD output signal
Negative
Level
37
OUT D4-3
―
――
BCD output signal
Negative
Level
*1: IN indicates signal input to the instrument, while OUT indicates signal output from the instrument.
*2: When BIN measurement is selected, refer to the top cell. When comparator measurement is selected, refer to the
middle cell. When BCD mode is selected, refer to the bottom cell.
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...