235
Discarding the Instrument
11.4 Discarding the Instrument
The instrument uses a lithium-ion battery as a power supply for saving the measurement condi
-
tions. When disposing of this instrument, remove the lithium battery and dispose of battery and
instrument in accordance with local regulations.
WARNING
•
To avoid electric shock, turn off the power switch and disconnect the power
cord and measurement cable, probe, or fixture before removing the lithium bat
-
tery.
•
Battery may explode if mistreated. Do not short-circuit, recharge, disassemble
or dispose of in fire.
•
Keep batteries away from children to prevent accidental swallowing.
Lithium Battery Removal
Required tools: One Philips screwdriver (No.1), One tweezers (to remove the lithium battery)
1
Verify that the power is off, and remove
the connection cables and power cord.
2
Remove the eight screws from the sides
and one screw from the rear.
Lithium bat-
tery
Battery
holder
3
Remove the cover.
4
Insert the tweezers between the battery
and battery holder as shown in the dia-
gram below and lift up the battery.
CAUTION
Take care not to short the + and -.
Doing so may cause sparks.
CALIFORNIA, USA ONLY
This product contains a CR Coin Lithium Battery which contains Perchlorate Material - special
handling may apply.
See www.dtsc.ca.gov/hazardouswaste/perchlorate
Maintenance and Service
11
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...