191
General Specifications
DC resistance measurement
Function
Measures DC resistance (when the measurement parameter is set to Rdc).
The measurement conditions are set separately from those for AC measurement.
(Measurement conditions: Measurement range, measurement speed, average,
judgment synchronization setting, DC delay, adjustment delay, line frequency)
Measurement signal level
Fixed at 1 V
Generating accuracy
±10% of setting ±20 mV
DC bias function
Function
Allows superposition of a DC voltage for measurement.
Operation mode
OFF/ON
Generating range
DC voltage : 0 V to 2.50 V (10 mV resolution)
When low Z high accuracy mode: 0 V to 1 V (10 mV resolution)
Generating accuracy
±10% of setting ± (V
AC
×0.01+30 mV)
V
AC
refers to the AC signal voltage setting (V).
Default setting
OFF
Residual charge protection function
Function
Protects the instrument from discharge voltage in the event that it is con
-
nected to a charged capacitor.
•
Discharge voltage reference value:
At 400 V DC or less, 5 J or less; at 1,000 V DC or less, 0.5 J or less
•
The energy W [J] stored in capacitance C [F] at voltage V [V] can be deter
-
mined using the following equation:
•
See "Appx. 6 The Residual Charge Protection Function" (p. Appx.9)
Terminal structure
4-terminal pair structure
Backup battery life
Approx. 10 years (at 25°C)
Used to back up the clock and settings (lithium battery).
Accuracy
Basic accuracy
Z: ±0.05% rdg.,
θ
: ±0.03° (representative value)
Guaranteed accuracy period
1 year
Warm-up time
60 minutes
Temperature and humidity
for guaranteed accuracy
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
Measurement time
Approx. 1 ms (measurement frequency: 1 MHz; measurement speed: FAST; no screen display)
Measurement speed
FAST, MED, SLOW, SLOW2 (Default setting: MED)
10
Specifications
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...