Performing Measurements in
LCR Mode
The LCR mode allows you to measure the impedance, phase angle, and other items by applying
any frequency or level (effective value) signal to the element you want to measure. This function is
suitable for evaluating the passive element of a capacitor, coil, or the like.
First, set the measurement mode to LCR mode (p.
3.1 Setting Display Parameters
You can select up to 4 of the 16 measurement parameters to display on the measurement screen.
These parameters are set on the measurement screen.
<Example> No. 1 parameter: Cs, No. 3 parameter: D (See "Parameters" (p. 40).)
1
Touch the No. 1 parameter key.
2
Touch the
Cs
key and then the
EXIT
key
to accept the settings.
1
2
3
Touch the No. 3 parameter key.
4
Touch the
D
key and then the
EXIT
key
to accept the setting.
1
2
Cs and D are set as the parameters.
If
OFF
is selected in the parameter setting, a
measurement value is not displayed.
3
Performing Measurements in LCR Mode
3
Performing Measurements in LCR Mode
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...