64
Setting Measurement Conditions (basic settings)
External trigger input method
There are the following three types of input method for a trigger.
•Touching the
TRIG
key on the screen to manually apply the trigger causes the instrument to perform one
measurement.
Measurement screen
SET
screen
•
Input via EXT I/O: Measurement is performed once each time a negative logic pulse signal is applied. See
"9.1 External Input/Output Connector and Signals" (p. 162).
•
Input from interface: Measurement is performed once when
*TRG
is transmitted.
Refer to the Communications Instruction Manual on the bundled LCR Application Disc.
Trigger delay (inserting a delay between the trigger and measurement)(Common)
The delay time period from input of the trigger signal to measurement can be set.
With this function it is possible to ensure that testing is started after the connection condition of the
object being tested and the test cables has stabilized.
The setting applies to both AC and DC measurement.
See "Trigger delay and Trigger synchronous output" (p. 65).
For more information about trigger delay timing, see the figures in "Measurement and data acquisi
Screen display method (For more information, see p. 24.):
(Measurement screen)
SET
key>(
SET
screen)
BASIC
tab>
DELAY
key
1
Use the
key to enter the trigger
delay time.
Settable range: 0 s to 9.9999 s with resolu
-
tion of 0.1 ms
When you want to turn off the trigger delay
setting, touch the
C
key.
(The set time is set to 0 s.)
2
Touch the
EXIT
key twice.
Displays the measurement screen.
When a trigger delay has been set, the measurement LED will light up from the time trigger input is received
until measurement completes.
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...