66
Setting Measurement Conditions (basic settings)
Measurement and data acquisition timing
Measurement and data acquisition timing varies with the following settings:
Trigger synchronous output (p. 65), Trigger delay (p. 64), Trigger synchronous delay (p. 65),
DC delay (p. 62), Adjustment delay (p. 63)
When the trigger synchronization function is on
Trigger
synchronous
delay
Data acquisition
AC measurement
(trigger synchronization on)
DC
delay
Adjustment
delay
Data
acquisition
Data
acquisition
DC measurement
(trigger synchronization on)
(DC adjustment on
)
Trigger
signal
DC measurement
(trigger synchronization on)
(DC adjustment off)
DC
delay
Data
acquisition
Trigger delay
Trigger delay
Trigger delay
DC measurement
after AC measurement
(trigger synchronization on)
(DC adjustment on
)
DC measurement
after AC measurement
(trigger synchronization on)
(DC adjustment off
)
Data acquisition
Switching time
(3.5 ms)
Data
acquisition
Adjustment
delay
Trigger
synchronous
delay
Trigger delay
Data
acquisition
Switching time
(
3.5 ms)
DC
delay
Data
acquisition
Trigger delay
Switching time
1.5 ms
Switching time
1.5 ms
DC
delay
Data
acquisition
Switching time
0.5 ms
Trigger
synchronous
delay
Trigger
synchronous
delay
Trigger
synchronous
delay
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...