63
Setting Measurement Conditions (basic settings)
Adjustment delay (setting the offset measurement delay time) (DC)
This delay time serves to delay measurement until offset measurement (0 V DC) stabilizes.
For more information about adjustment delay timing, see the figures in "Measurement and data
Screen display method (For more information, see p. 24.):
(Measurement screen)
SET
key>(
SET
screen)
Rdc
tab>
ADJ DELAY
key
1
Change the adjustment delay time with
the
keys.
Settable range: 0.0030 s to 9.9999 s
To revert the setting to the default value,
touch the
C
key.
(The set time is set to 0.0030 s.)
2
Touch the
EXIT
key twice.
Displays the measurement screen.
The time required until the DC signal level stabilizes differs depending on the test sample to be measured.
To ensure measurement is performed accurately, observe the measurement waveform in advance and then
set the delay time required until the DC signal level stabilizes.
Trigger (perform measurements with user-defined timing) (Common)
Trigger functionality allows you to start and stop recording based on a particular signal.
When recording is started or stopped by a specific signal, we say the trigger is "gapplied" or "trig
-
gering occurs".
With this instrument, you can select the following two types of trigger.
Internal
Trigger
Generates the trigger signal internally and repeats measurement automatically.
External
trigger
Accepts input of signals and commands from an external source to control measure
-
ment. You can also apply the trigger manually on the instrument’s screen.
This setting applies to both AC and DC measurement.
Screen display method (For more information, see p. 24.):
(Measurement screen)
SET
key>(
SET
screen)
BASIC
tab>
TRIG
key
1
Select the trigger type.
INT
Internal trigger
Automatically repeats measure-
ment.
EXT
External trigger
Input the trigger manually via EXT
I/O or from the interface.
2
Touch the
EXIT
key twice.
Displays the measurement screen.
3
Performing Measurements in LCR Mode
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...