46
Setting Measurement Conditions (basic settings)
Setting AUTO ranging
Screen display method (For more information, see p. 24.):
AC measurement: (Measurement screen)
SET
key>(
SET
screen)
BASIC
tab>
RANGE
key
DC measurement: (Measurement screen)
SET
key>(
SET
screen)
Rdc
tab>
RANGE
key
1
Touch the
AUTO
key.
2
Touch the
EXIT
key twice.
Displays the measurement screen.
•
If the instrument is being used outside the
limits of its specification, the suitable range
may not be set in auto ranging function.
Check the accuracy assured ranges in
"Guaranteed accuracy measurement level
range" (p. 213) and then change the test
conditions.
•Manually changing the set range while us-
ing the
AUTO
setting will cause the instru-
ment to switch to the
HOLD
setting.
The AUTO range limit function allows you to limit the AUTO ranging range.
1
Touch the
MIN
key.
2
Select the AUTO range lower limit range.
3
Touch the
EXIT
key.
The display will return to the screen shown
in step 1.
4
Touch the
MAX
key, and select the
AUTO range upper limit range.
5
Touch the
EXIT
key twice.
Displays the measurement screen.
The AUTO range scope is the selection of ranges
within which the AUTO range is selected. If the
AUTO range scope has been limited, the instrument
will not select a range that lies outside that scope.
For more information about the AUTO range scope,
see "Measurement range" (p. 190).
•
When canceling the AUTO range limit function, set the lower
limit range to 100 m
Ω
and the upper limit range to 100 M
Ω
.
Range selection screen when the AUTO range scope has
been limited
Example: When the lower limit range is set to 1 k
Ω
and the up-
per limit range is set to 1 M
Ω
Ranges that fall outside the AUTO
range scope will not be shown.
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...