Setting Measurement Conditions (basic settings)
Averaging (limiting display value instability) (AC/DC)
With the averaging function, the measured values can be averaged. This function can be used to
limit instability of displayed measured values.
AC measurement
With internal trigger
A rolling average of the tested values over the set number of times for
averaging is always calculated backwards from the present.
(When the sample to be tested is changed over, it takes a little time for a
certain stabilization time period until the results is reliable.)
With external trigger
Average over the number of averaging times based on trigger input.
DC measurement
Averaging processing during DC measurement calculated an arithmetic average regardless of the
trigger setting.
Screen display method (For more information, see p. 24.):
AC measurement (Measurement screen)
SET
key>(
SET
screen)
BASIC
tab>
AVG
key
DC measurement (Measurement screen)
SET
key>(
SET
screen)
Rdc
tab>
AVG
key
1
Use
key to enter the number of
averaging times.
Settable range: 1 to 256
To disable the averaging function, touch
the
C
key.
(The setting will be set to 001.)
2
Touch the
EXIT
key twice.
Displays the measurement screen.
When the range is changed, including by
auto-range operation, averaging processing
up to that point is cleared and then resumed.
3
Performing Measurements in LCR Mode
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...