50
Setting Measurement Conditions (basic settings)
1
Select the measurement signal mode.
V
Open circuit voltage (V) mode
CV
Constant voltage (CV) mode
CC
Constant current (CC) mode
2
Use the
key to enter voltage level
or current level.
Measurement
signal mode
Settable range
V, CV
4 Hz to 1.0000 MHz:
0.010 V to 5.000 V
1.0001 MHz to 8 MHz:
0.010 V to 1.000 V
CC
4 Hz to 1.0000 MHz:
0.01 mA to 50.00 mA
1.0001 MHz to 8 MHz:
0.01 mA to 10.00 mA
3
Touch the
EXIT
key twice.
Displays the measurement screen.
•
When low Z high accuracy mode (p. 56)
is on, the valid setting range will vary.
Measurement
signal mode
Settable range
V, CV
0.010 V to 1.000 V
CC
•When the output im-
pedance is 10
Ω
:
0.01 mA to 100.00 mA
•When the output im-
pedance is 100
Ω
:
0.01 mA to 10.00 mA
See: "For setting range and accuracy"
•
The accuracy of testing varies according to
the test signal level.