56
Setting Measurement Conditions (basic settings)
Low Z High Accuracy Mode (high-precision measurement) (AC/DC)
Enabling low Z high accuracy mode will switch the output resistance to 10
Ω
, enabling high-preci
-
sion measurement by allowing an adequate amount of current to flow to the measurement sample.
Screen display method (For more information, see p. 24.):
AC measurement (Measurement screen)
SET
key>(
SET
screen)
BASIC
tab>
RANGE
key
DC measurement (Measurement screen)
SET
key>(
SET
screen)
Rdc
tab>
RANGE
key
1
Touch the
LOW Z ON
key.
2
Touch the
EXIT
key twice.
Displays the measurement screen.
•
In low Z high accuracy mode, the measure
-
ment signal level setting range changes.
•Changing the low Z high accuracy mode
setting while open correction, short correc
-
tion, or load correction is enabled causes
the correction values to be disabled.
•Low Z high accuracy mode can only be
enabled while using the 100 m
Ω
, 1
Ω
, or 10
Ω
range. See the table below.
Measurement
range
DC measure-
ment
AC measurement (measurement frequency)
to 1 kHz to 10 kHz to 100 kHz to 1 MHz to 5 MHz to 8 MHz
100 M
Ω
10 M
Ω
None
1 M
Ω
100 k
Ω
10 k
Ω
Even if low Z high accuracy mode is set to
ON
, the output resistance will remain 100
Ω
.
(The measurement signal level will be limited to 1 V or less.)
1 k
Ω
100
Ω
10
Ω
Low Z high accuracy mode is enabled.
(When low Z high accuracy mode is off, the output resistance will
be 100
Ω
.)
1
Ω
100 m
Ω
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...