166
External Input/Output Connector and Signals
Pin
I/O
*
Signal name
Function
Logic
19
-
(Unused)
-
-
-
20
-
(Unused)
-
-
-
21
-
(Unused)
-
-
-
22
-
(Unused)
-
-
-
23
-
(Unused)
-
-
-
24
-
(Unused)
-
-
-
25
-
(Unused)
-
-
-
26
-
(Unused)
-
-
-
27
-
ISO_COM
Isolated common signal ground
-
-
28
OUT EOM
―――
This signal indicates that measurement is complete. “HI
(OFF)” indicates that measurement is in progress, while
“LO (ON)” indicates that measurement is complete. When
LO (ON), the comparator judgment results have been
finalized.
Falling
Edge
29
OUT
INDEX
―
―――
―
Signal indicating that A/D conversion for the measurement
circuit has completed: When this signal changes from HI
(OFF) to LO (ON), the sample may be changed.
Falling
Edge
30
OUT
No.1_PARA1-IN
―
―― ―
―――
――
―
Outputs IN comparator judgment results for the first No. 1
parameter.
Negative
Level
31
OUT
No.1_PARA3-HI
―
―― ―
―――
――
―
Outputs HI comparator judgment results for the first No. 3
parameter.
Negative
Level
32
OUT
No.1_PARA3-LO
―
―― ―
―――
――
―
Outputs LO comparator judgment results for the first No. 3
parameter.
Negative
Level
33
OUT
No.2_PARA1-HI
―
―― ―
―――
――
―
Outputs HI comparator judgment results for the second
No. 1 parameter.
Negative
Level
34
OUT
No.2_PARA1-LO
―
―― ―
―――
――
―
Outputs LO comparator judgment results for the second
No. 1 parameter.
Negative
Level
35
OUT
No.2_PARA3-IN
―
―― ―
―――
――
―
Outputs IN comparator judgment results for the second
No. 3 parameter.
Negative
Level
36
-
(Unused)
-
-
-
37
-
(Unused)
-
-
-
*: IN indicates signal input to the instrument, while OUT indicates signal output from the instrument.
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...