6
Measurement process
Measurement process
This section uses AC measurement of a laminated ceramic capacitor as an example to provide an
overview of the instrument’s functionality.
Items to be prepared:
9263 SMD Test Fixture, Laminated ceramic capacity you want to measure
1
Inspect the instrument before measurement. (p. 33)
2
Connect the power cord to the instrument. (p. 34)
3
Turn on the power. (p. 36)
(A 60-minute warm-up period is necessary before performing the correction process de
-
scribed in Step 9.)
4
Set the date and time. (p. 38)
5
Connect the 9263 SMD test fixture to the measurement terminals.
•Overview of connections: p. 35
(The connection method varies with the probes and
fixture being used. For more information, see each prod
-
uct’s user manual.)
•
Optional probes and fixtures: p. 2
6
Set the measurement mode to LCR. (Default setting: LCR)
2
3
1
Use the
CONTINUOUS
setting if you wish to take continuous measurements under multiple sets of
conditions. (In LCR mode, you must first set and save the measurement conditions.)
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...