33
Pre-Operation Inspection
2.2 Pre-Operation Inspection
Please read the “Operating Precautions” (p. 12) before use.
Before using the instrument, verify that it operates normally to ensure that no damage occurred
during storage or shipping. If you find any damage, contact your authorized Hioki distributor or
reseller.
1
Peripheral Device Inspection
Is the power cord insulation torn,
or is any metal exposed?
No Metal Exposed
Metal
Exposed
Do not use the instrument if damage is found,
as electric shock or shortcircuit accidents could
result. Contact your authorized Hioki distributor
or reseller.
Is the insulation on a cables, or
is any metal exposed?
No Metal Exposed
Metal
Exposed
If there is any damage, measured values may
be unstable and measurement errors may occur.
Replace the cable with an undamaged one.
2
Instrument Inspection
Is damage to the instrument evi
-
dent?
Yes
If damage is evident, request repairs.
No
When turning power on
Does the Opening screen appear
(model no., version no.)?
No
The power cord may be damaged, or the instru
-
ment may be damaged internally.
Request repairs.
“11.2 Troubleshooting” (p. 223)
Yes
Is there an error display on the
Opening screen?
No
An error
indication
occurs
(Err)
The instrument may be damaged
internally. Request repairs.
See “Transporting the instrument” (p. 222).
Inspection complete
2
Measurement Preparations
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...