70
Judging Measurement Results
Setting the judgment mode
Select a judgment mode as described below and configure the settings.
Screen display method (For more information, see p. 24):
(Measurement screen)
SET
key>(
SET
screen)
ADVANCED
tab>
JUDGE
key
1
Select the judgment mode.
OFF
Disables comparator and BIN func
-
tion.
COMP
Enables comparator function.
Configure the comparator function
BIN
Enables BIN function. Configure the
BIN function settings (p. 75).
2
Touch the
EXIT
key twice.
Displays the measurement screen.
•
When comparator measurement and BIN measurement are performed, only the first and third parameters
can be set. (During BIN measurement, the No. 3 parameter will be displayed in the No. 2 parameter area.)
•
During comparator measurement, the No. 2 and No. 4 parameter displays will indicate [
LMT
].
•
During BIN measurement, the No. 4 parameter display will indicate [
BIN
].
Configuring comparator function settings (judging measurement re
-
sults based on one judgment standard)
The comparator function allows you to do the following.
Preset a reference value and upper and lower limit values
as the judgment reference, and display a judgment result
as
HI
(higher than the upper limit value),
IN
(within the
range set for the upper and lower limit values), or
LO
(lower
than the lower limit value).
• Output the judgment results to an external device (via the
EXT I/O connector).
•
Select different settings and perform judgment for up to
two parameters.
•
Be notified of judgment results by buzzer.
See "Key tones and judgment tones" (p. 90).
•
Confirm the judgment result from the judgment result indi
-
cation LEDs on the front panel of the instrument.
When the compar-
ator measurement
result is
IN
, the
green indicator
will light up.
When the compara-
tor measurement
result is
HI
or
LO
,
the red indicator will
light up.
(Judgment result indication LEDs)
HI
Measured value is above upper limit
IN
Upper limit value ≥ calculated value ≥
lower limit value
LO
Measured value is below lower limit
- - -
When no reference standards have been
set
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...