75
Judging Measurement Results
Configuring BIN function settings (judging measured values based on
multiple judgment standards)
Set the upper and lower limit values for two parameters and display up to 10 classifications of judg
-
ment results. You can also output the judgment results to an external device.
When the measurement result
falls within the BIN category
range, the green indicator will
light up.
When the measurement
result is
OUT OF BINS
, the
red indicator will light up.
(Judgment result indication LEDs)
BIN2 (numeral)
When BIN judgment
- - -
When BIN is not set
OUT OF BINS
When does not match any BIN
(Measurement screen)
About BIN measurement
Bin measurement is performed using a process such as the example shown below.
Judgment BIN
number
No. 1
Parameter
No. 3
Parameter
BIN Judgment
results
Judgment
Judgment
Judgment
Judgment
Judgment
Judgment
Judgment
Judgment
The instrument will display the first BIN number for which the measurement value is judged to be
within the set judgment standard.
If none of the BIN judgments is determined to apply,
OUT OF BINS
will be displayed.
The BIN number for which the measured
value was first judged to be within the set
judgment standards will be displayed as the
judgment result.
Judgment reference
3
Performing Measurements in LCR Mode
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...