189
General Specifications
Limit function
Function
Sets limits (limit values) on the voltage that can be applied to the sample or
the current that can flow to the sample.
Operating mode
OFF/ON
Current limit
During open-terminal voltage mode or constant-voltage mode operation
Limit range: 0.01 mA to 100.00 mA
Limit accuracy: Monitor current accuracy specifications and software control
range (±1%±10 µA)
Voltage limit
During constant-current mode operation
Limit range: 0.01 V to 5.000 V
Limit accuracy: Monitor current accuracy specifications and software control
range (±1%±10 mV)
Default setting
OFF
Output impedance (H
CUR
terminal, measurement frequency 1 kHz)
Normal mode
100
Ω
±10
Ω
*
Low impedance high accu-
racy mode
10
Ω
±2
Ω
*: The output resistance and termination resistance will be set to 50
Ω
(The output resistance is the resistance
connected to the H
CUR
terminal, and the termination resistance is the resistance connected to the L
CUR
termi-
nal) when measuring under the following conditions with the cable length set to 1 m, 2 m, or 4 m:
All measurement frequencies for the 10 k
Ω
range to the 100 M
Ω
range
Measurement frequencies from 1.0001 MHz to 8 MHz for the 100 m
Ω
range to the 1 k
Ω
range
10
Specifications
Содержание IM3536
Страница 20: ...16 Operating Precautions ...
Страница 34: ...30 Screen Layout and Operation ...
Страница 140: ...136 Testing the System Self diagnosis ...
Страница 224: ...220 About Measurement Times and Measurement Speed ...
Страница 240: ...236 Discarding the Instrument ...
Страница 254: ...Appx 14 Dimensional Diagram Appx 10 Dimensional Diagram 230 114 15 48 33 33 25 35 119 1 330 1 Unit mm ...
Страница 264: ...Appx 24 Device Compliance Statement ...
Страница 270: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...